ISO-DATA
"20/20" SERIES

20/20 Series Multi-Well Gamma Counters (refurbished)
"The Gamma Counting System That Grows With Your Needs"
10 or 20
Multi-detector Counting Systems
The Model 20-10 has ten detectors, and the Model 20-20 has twenty
detectors. Model 20-10 may be upgraded to twenty detectors as future
needs require. Each model has variable windows. Two isotopes can be
counted simultaneously. This Real Time data reduction system is
extremely versatile, allowing the user a choice in processing data.
The micro-computer offers the ultimate in data handling. Quality
Control parameters may be chosen and stored for the various assays.
Histograms, Levey-Jennings plots and other data may be printed. It
is a very versatile RIA and quality control system.
Model 20/20 Series Gamma Counter Specifications:
Highest Throughput
This model is capable of counting 100 samples in less than 7 minutes.
A traditional gamma counter would take nearly 2 hours to count the same
group of samples.
Dual Label
Dual label assays are handled easily, since these counters allow the
counting of two isotopes simultaneously.
Spilldown Correction
When counting dual label samples, there is usually a presence of the
high energy isotope in the low energy window. This high energy,
present in the low energy window, is automatically subtracted with
spilldown correction.
Variable
Windows
Both channel A and Channel B windows are user adjustable.
Normalization of Detectors
With the normalization program, all detectors can be matched to assure
you of reproducible results year after year.
Disable Well
If for some reason, one or more detectors cannot be normalized, that
detector can be taken out of service without influencing replicate
samples. Since there are multiple detectors, there is no need to
discontinue counting samples.
Accepts a Variety of Tube Sizes
We realize that all RIA tests are not done in 12x75mm tubes. The
20/20 Series Gamma Counters have the flexibility to accommodate many
different size test tubes.
Counted Tube Storage
The 20/20 series counter will store up to 600 counted tubes for replay
with various curves.
Display
LED display provides counts for each detector for both Channel A and
Channel B, time and tray number
Dual External Ports
One port is for the terminal, the second port has been provided for
interfacing with a central lab computer.
|
Detectors |
Nal (TI) integral line thin wall well crystal |
|
Detector Efficiency |
Better than 65% for 125-I |
|
Background |
Less than 100 CPM for 125-I |
|
Count Rate Linearity |
High speed electronics and automatic dead time
correction allow better than 2% for total count rates up to 10-6
CPM |
|
Temperature Stability |
Less than 5% count rate variation for 125-I
window across operational range of +10°C to +40°C |
|
Voltage Stability |
Less than 1% count rate variation for 125-I
window at nominal line voltage +/- 15% |
|
Storage Temperature |
-18°C to +50°C |
|
Humidity |
Up to 90% RH non-condensing |
|
Tube Size |
Up to 15mm diameter length unlimited |
|
Shielding |
Virgin Lead |
|
Visual Display |
LED, 8 digits |
|
Printer |
Full Page, Parallel |
|
Dimensions |
24" wide, 19" deep, 16" high |
|
Weight |
10 well - 105 lbs.
20 well - 175 lbs. |
|
Power |
10 well - less than 150 VA
20 well - less than 250 VA |
Call 1-800-722-8072
or
email for more information!